BoilerSpection IM Manual Troubleshooting 29
6.6 Air Filtration System
Symptom Probable Cause Corrective Action
Loss of air
pressure
Clogged filter Check that filters are not
clogged up and follow proper
maintenance schedule (see
section 5)
Leaky connections Check that the connections
are secure and tighten as
necessary
Leaky/damaged hose Contact LumaSense (refer to
section 1.7)