Pearl Control Software Manual Using Calibrate Tab Features 29
2. Click on the Calibrate Tab.
3. Click on Gain Map.
4. Point the camera at a cold target and click Set Low Point.
Note: If no low point is set, the shutter is used as a reference point.
4.3.4 Set High Point
The Set High Point sets the high point in the gain map calculation. The High Point calculation
is always performed after the Low Point is set.
To set the high point:
1. Click on the Calibrate Tab.
2. Click on Gain Map.
3. Point the camera at a hot target and click Set High Point.
4.3.5 Save Map
Save Map saves the gain map to the camera permanently. This operation will take a few seconds
to perform and the camera will flicker during saving.
To save the gain map to the camera permanently:
1. Click on the Calibrate Tab.
2. Click on Gain Map.
3. Click Save Map.