Strona 1 - IMPAC Pyrometer

MANUALIMPAC PyrometerISR 320

Strona 2

ISR 320 Manual Introduction  10OpticsSighting:Built-in LEDtargeting light andLEDs for intensity alignmentOptics: Fixed optics a=300 mm or a=800 mmDi

Strona 3 - Contents

ISR 320 Manual Introduction  112.4 Dimensions2.5 Physical User Interface2.6 Accessories (optional)Numerous accessories guarantee easy installation o

Strona 4 - ISR 320 Manual Contents  iv

ISR 320 Manual Introduction  122.6.3 Air PurgeThe air purge protects the lens from contamination of dust andmoisture. It has to be supplied with dry

Strona 5 - 1 General Information

ISR 320 Manual Controls and Installation  133 Controls and Installation3.1 Electrical InstallationThe pyrometer is powered by 24 V DC ± 25% (very we

Strona 6 - 1.4 Unpacking the Instrument

ISR 320 Manual Controls and Installation  143.1.2 Connecting the pyrometer to a PCThe pyrometer is equipped with an RS485 serial interface. With the

Strona 7

ISR 320 Manual Controls and Installation  15can also be connected with its serial interface, whereas the digital display DA 6000-N has to beconnecte

Strona 8

ISR 320 Manual Controls and Installation  163.3 Optics3.3.1 Spot SizesThe ISR 320 has fixed optics for 300 mm or 800 mm measuring distance.The table

Strona 9 - 2 Introduction

ISR 320 Manual Settings / parameter descriptions  174 Settings / parameter descriptionsThe pyrometer is equipped with a wide range of settings for o

Strona 10

ISR 320 Manual Settings / parameter descriptions  18Note: Emissivity Slope K Settings: 0.800 to 1.250 in steps of 1/1000 (2-color mode).4.2.1 Temper

Strona 11 - 2.6 Accessories (optional)

ISR 320 Manual Settings / parameter descriptions  194.3 Transmittance Transmittance is a parameter that can compensate for signal loss due to exter

Strona 12 - 2.6.6 Flange System

Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us

Strona 13 - 3 Controls and Installation

ISR 320 Manual Settings / parameter descriptions  20Double StorageMode:Double storage mode comes into effect when selecting one of the resetinterval

Strona 14 - 3.1.3 Connection to RS485

ISR 320 Manual Settings / parameter descriptions  214.7 Relative Signal StrengthRelative signal strength stands for the product of emissivity, surfa

Strona 15 - 3.2 Sighting

ISR 320 Manual Settings / parameter descriptions  224.11 Sub RangeYou have the opportunity to choose a sub range (minimum span 50 °C) within the bas

Strona 16 - 3.3 Optics

ISR 320 Manual Settings / parameter descriptions  23Close below:The contact closes (switches to ground), if the temperature falls below theentered v

Strona 17 - 4.2 Emissivity Slope K

ISR 320 Manual Settings / parameter descriptions  24To ensure consistent document formatting, this page was intentionally left blank.

Strona 18

ISR 320 Manual Software InfraWin  255 Software InfraWinThe operating and analyzing InfraWin software is included with delivery of the pyrometer. Ina

Strona 19 - 4.5 Clear Peak Memory (t

ISR 320 Manual Software InfraWin  265.4 Basic settingsAll preset values for thedevice can be displayed andmodified, if necessary under theDevices/Pa

Strona 20 - 4.6 Analog Output

ISR 320 Manual Software InfraWin  27In the lower part of the window, the connection with the preset baud rate can be checked. Herethe command was se

Strona 21 - 4.10 Operating Modes

ISR 320 Manual Software InfraWin  285.6 Warning Level / Switch-Off LevelIn addition to the relative signal bar graph, the warning level and the swit

Strona 22 - 4.15 Limit Switch (SP1)

ISR 320 Manual Software InfraWin  29Once the temperature entry has beenentered and confirmed with OK,InfraWin will then calculate theemissivity slop

Strona 23

ISR 320 Manual Contents  iiiContents1 General Information ...

Strona 24

ISR 320 Manual Software InfraWin  30frequency that readings were taken (settings at 5.9 PC sampling rates). As the amount of dataincreases, so does

Strona 25 - 5 Software InfraWin

ISR 320 Manual Software InfraWin  31After entering the aperture and the main spot size, the input of interim values calculates spotsizes in differen

Strona 26 - 5.4 Basic settings

ISR 320 Manual Software InfraWin  32To ensure consistent document formatting, this page was intentionally left blank.

Strona 27 - 5.5 Relative Signal

ISR 320 Manual Data format UPP 336 Maintenance6.1 Cleaning ISR 320 windowBecause there are no moving parts in the ISR 320, the only regular mainten

Strona 28

ISR 320 Manual Data format UPP 34In the 1-color (mono) mode, the pyrometer can measure objects at any distance.However, the object has to be bigger

Strona 29 - 5.9 Listing (analyzing)

ISR 320 Manual Data format UPP 357 Data format UPP (Universal PyrometerProtocol)Software commands can be exchanged directly with the pyrometer thro

Strona 30 - 5.13 Spot size calculator

ISR 320 Manual Data format UPP 36Description Command ParametersMeasuring value(one-channel andratio temperature)AAek Answer: SSSSSQQQQQ2x5 decimal

Strona 31 - 5.14 Search I/O Module

ISR 320 Manual Data format UPP 37Description Command ParametersMeasuring value AAms Answer: QQQQQ (88880=Overflow)5 decimal digit (in °C or °F, la

Strona 32

ISR 320 Manual Data format UPP 38To ensure consistent document formatting, this page was intentionally left blank.

Strona 33 - 6 Maintenance

ISR 320 Manual Reference Numbers  398 Reference Numbers8.1 Reference numbers instrumentTemperature Range Reference Number a / mmISR 320, 700 to 1700

Strona 34

ISR 320 Manual Contents  iv4.4 Response Time (t90) ...

Strona 35 - Protocol)

ISR 320 Manual Reference Numbers  403 852 290Power supply NG DC for DIN rail mounting; 100 to 240 V AC  24 V DC, 1 A3 852 550Power supply NG 2D for

Strona 36 - =0.100 to 1.000

ISR 320 Manual Troubleshooting  419 TroubleshootingSymptom Probable Cause CommentsNo analog output even if thedisplay shows a temperatureabove lower

Strona 37

ISR 320 Manual Troubleshooting  42Symptom Probable Cause CommentsIncorrect slope adjustments. See sections 4.2, 5The measurement is influenced by re

Strona 38

ISR 320 Manual Troubleshooting  43Symptom Probable Cause CommentsNoisy readings (fastfluctuations).Improper grounding of cable shield and/orISR 320

Strona 39 - 8 Reference Numbers

ISR 320 Manual Troubleshooting  44To ensure consistent document formatting, this page was intentionally left blank.

Strona 40

ISR 320 Manual Index  45Index11 measure… 26AAccesories 11Air Purge 12Analog Output 20Appropriate use 9BBasic settings 26Baud Rate 22CCalibration 33L

Strona 41 - 9 Troubleshooting

ISR 320 Manual  46SSafety vScaling Trend button 29Scope of delivery 9Service Request 6Settings / parameter descriptions 17Sighting 15Single Storage

Strona 42

ISR 320 Manual Contents  v1 General Information1.1 Information about the user manualCongratulations on choosing the high quality and highly efficien

Strona 43

ISR 320 Manual General Information  61.3 Limit of liability and warrantyAll general information and notes for handling, maintenance and cleaning of

Strona 44

ISR 320 Manual General Information  7Technical Support can be contacted by telephone or email:Santa Clara, California Telephone: +1 408 727 1600 or

Strona 45 - ISR 320 Manual Index  45

ISR 320 Manual General Information  8To ensure consistent document formatting, this page was intentionally left blank.

Strona 46 - ISR 320 Manual  46

ISR 320 Manual Introduction  92 Introduction2.1 Appropriate useThe IMPAC ISR 320 pyrometer is a stationary, digital pyrometer for non-contact temper